What is a light incubator?
The light incubator is a special constant temperature culture device for the cultivation and breeding of bacteria, molds and microorganisms. It is especially suitable for scientific research and production in the fields of bioengineering, medical research, agriculture and forestry, aquatic products, and animal husbandry. Today we will mainly introduce the use method and precautions of the light incubator, hoping to help users better apply the product.
How to use the light incubator:
Plug in the power supply, turn the power switch to the "ON" position, then the power indicator light is on, press the "power" button, all the digital display "888……" indicator lights are all on, after displaying 4S, the control enters the running state, this operation When the time is counted down, the display flashes in seconds when starting up; the "query" button displays the parameter setting value in the running period of this time, and then press the "query" button to display the actual parameter; long press the set button 4S to enter the 24-segment parameter setting Setting, the setting indicator light is on, and the time window displays "P-01" to indicate the first parameter setting (that is, the parameter in the first hour, that is, the 24:00-23:00 time period.) The light 1 window flashes, and it can be passed △/▽ key to modify the light setting value of the first segment (0-4)(0.0-60.0)℃, at this time, if you press the exit key, you can return to the light setting state. Press the set button again, the time window displays "P-02" to enter the second stage of parameter setting, cycle in turn, set 24 stages, each stage represents the set value of each hour. Press the set button to exit the setting state; when setting, try to make the setting state consistent with the natural environment; if you set in the morning, you can set the light and temperature values first, and when you enter the night according to the time period, set no Light, etc.; during operation, if you find that the temperature control accuracy is poor, long press the temperature key for 4S to enter the internal parameter state, adjust the parameters, and after the adjustment is completed, long press the temperature key to exit the parameter setting.
Matters needing attention in light incubator:
1. After the equipment has landed, if the ground is uneven, it should be leveled. Wipe the tank wall and the surface of the equipment frequently to keep it clean.
2. When the equipment is in normal operation, the placement of the load in the box should not affect the air circulation to ensure uniform temperature and humidity in the box.
3. The equipment should be moved in parallel, and the inclination angle in any direction should be less than 45 degrees.
4. If the equipment is not used for a long time, the power cord should be unplugged to prevent the equipment from being charged and hurting people. And it should be run regularly (usually a quarter) according to the conditions of use for 2-3 days to drive away the moisture of electrical components and avoid damage to related devices.
Welcome to the official website of Runhua Instruments. We are a professional manufacturer of laboratory instruments. Quality and reputation a…
The water bath constant temperature shaker is a biochemical instrument that combines a temperature controllable constant temperature water ba…
Series of biochemical incubators:Features of 150A biochemical incubator: the biochemical incubator has the function of cooling and heating tw…
High-precision stirring water bath produced by Runhua Instruments:1) The general water bath does not have the circulating stirring function. …
Maintenance and maintenance of electric mixer1. Correct use and attention to the maintenance of the instrument to keep it in good working con…
What work should be done in the use of digital display stainless steel electric heating plate? Digital display stainless steel electric h…