Operation and use of tissue crushing homogenizer

  1. First align the cutter shaft and the machine shaft, and then move the spring nature of the connector down into the groove, (moving the foot down and the step firmly) before the machine can operate.

  2. Before turning on, pay attention to whether the voltage on the power supply and the motor match.

  3. The carbon brush should be checked frequently. If it is too short, replace it with a new one. If the spark of the carbon brush is found to be abnormal, stop using it immediately, check the cause of the obstacle, and repair it before using it to prevent the rectifier and coil from burning out.

  4. Due to the high speed of the motor, continuous use will cause the motor to burn out, so the use time is limited to three minutes, and it can be used after stopping for 5 minutes.

  5. When the glass cup is installed, it needs to be aligned with the center of the circle on the machine, centered, and there is no shaking on all sides before it can be turned on.

  6. The machine should not be idling, a small amount of liquid or grease must be put in when using it.

  7. When putting in the material, it must be added slowly. First, start the slow gear and then the fast gear. However, the slow gear can only be used as a starting point and should not be used frequently.
Related articles