Instructions for use of quartz sub-boiling high-purity water distiller
Information sources: | Release date:
2019-07-05 14:31:12
| Browse volume:195
SYZ-550 Quartz Subboiling High Purity Water Distiller
user's manual
I. Overview
Ultra-high purity water prepared by quartz sub-boiling has not only low water ion content, but also large water output. The instrument is all made of high-quality quartz glass, and the distilled water does not contact any metal, so the water quality is extremely high.
2. Technical parameters
1. Voltage: 220V/50Hz
2. Power rate: 1500W
3. Water output: 800-1000mL/h
4. Instrument size:
Three, installation
1. Place the instrument on the workbench (close to the water source and power supply), and insert the frosted end of the condenser into the top of the frosted mouth of the horizontal flask.
2. The overflow hose and the condensate outlet hose can be introduced into the waste water tank; then the condensed water inlet hose is connected to the commercially available tap water pipe; then the water inlet hose is connected to the steamed raw water, and the raw water inlet hose is equipped with a water stop clamp , And finally connect the condenser interface hose to the receiving bottle.
Four, work
1. Slowly turn on the tap water switch, the water flows into the inlet of the condenser, and the water is discharged from the serpentine tube to the condenser tube. The water flow should not be too rapid to avoid damage to the instrument. Then add the steamed raw water and leave it at the overflow port until the water comes out, or use a water stop clamp to control the flow rate, and then it can work continuously.
2. Turn on the power switch, that is, the sub-boiling starts to work, and the water starts to produce after about 2 minutes, which is high-purity water.
Five, maintenance
1. At te end of each work or long-term use, the condensed water and distilled water should be drained out, and the distilled water should be drained from the water inlet. The condensed water can be discharged by unplugging the tap. This will prevent the accumulation of scale over time.
2. For long-term use of scaling, 50% HNO3 can be poured in from the distilled water inlet and immersed to the scaling place. After soaking for 2-4 hours, the hose can be drained by lowering the original water inlet. Then pour in distilled water or deionized water and rinse 3-5 times to reuse.
Six, cleaning
1. The new distiller needs to be used directly after introducing distilled water or deionized water. The quality of the water distilled in the first two days can be used for cleaning pipe walls and washing quartz bottles and polyethylene barrels. The contamination of the container cannot be ignored.
2. If the scale is serious or polluted by adding chemical reagents (such as potassium permanganate, etc.) for a long time, it can be soaked in 20% HF, but the time should not be too long, depending on the smoothness of the distiller, and deionize after soaking Rinse with water, balance and purify.
3. The distiller will not be used for a long time. All outlets should be sealed with filter paper to avoid environmental pollution. It can also be soaked in 2NHNO3 for storage.
4. Our factory pays special attention to after-sales service. If you find other unfinished matters in the future use, please feel free to contact our factory.
Related articles
-
2020-07-21 11:05:15
Welcome to the official website of Runhua Instruments. We are a professional manufacturer of laboratory instruments. Quality and reputation a…
-
2020-06-10 10:18:30
The water bath constant temperature shaker is a biochemical instrument that combines a temperature controllable constant temperature water ba…
-
2020-06-09 09:31:04
Series of biochemical incubators:Features of 150A biochemical incubator: the biochemical incubator has the function of cooling and heating tw…
-
2020-06-01 13:28:13
High-precision stirring water bath produced by Runhua Instruments:1) The general water bath does not have the circulating stirring function. …
-
2020-06-01 13:24:47
Maintenance and maintenance of electric mixer1. Correct use and attention to the maintenance of the instrument to keep it in good working con…
-
2020-05-29 15:47:15
What work should be done in the use of digital display stainless steel electric heating plate? Digital display stainless steel electric h…
Related products