Welcome to visit Changzhou Runhua Electric Co., Ltd. Official website!
TEL

0086-519-82305398

Current location:Home > Products > Incubator series > Mould Cultivation Cabinet
PRODUCTS
  • Microcomputer intelligent temperature and humidity control in mould incubator

Microcomputer intelligent temperature and humidity control in mould incubator

Overview:1、 Function and application of mould incubator: Mold incubator is suitable for environmental protection, health and epidemic prevention, agriculture, forestry, animal husbandry, aquaculture and other scientific research and production departments. It…

Consulting

1、 Function and application of mould incubator:

  Mold incubator is suitable for environmental protection, health and epidemic prevention, agriculture, forestry, animal husbandry, aquaculture and other scientific research and production departments. It is a necessary test equipment for water analysis, bacteria, mold, microbial culture, plant cultivation, breeding test.


2、 Technical parameters of mould incubator:

1. The temperature control system of this machine adopts microcomputer single chip microcomputer technology, which has high temperature control precision;

2. With temperature and time adjustment control, touch keyboard setting adjustment;

3. Display for double screen high brightness digital tube display, accurate and intuitive display value, superior performance;

4. It has super ultraviolet sterilization function to ensure pure culture environment;

5. The body shell is sprayed with plastic, which makes the whole machine beautiful and elegant;

6. The inner tank is made of anti-corrosion and anti rust materials; the box body is treated with foaming to ensure the quality of heat preservation and refrigeration

7. The studio is equipped with lighting device, large viewing angle, vacuum tempered glass, easy to observe;

8. The height and number of shelves in the studio can be adjusted according to the user's requirements

9. Mold UV germicidal lamp is equipped with a special switch, which can be turned on when necessary. After turning on the lamp, please do not look directly at it to avoid damaging your eyes. When you do not need to use the sterilization lamp, try not to turn it on as far as possible, because the ultraviolet light is very strong, in order to prevent you from being hurt.



3、 Technical parameters of mould incubator:

Model: shp-50

Refrigeration mode: stainless steel heater, totally enclosed compressor

Volume: 50 liters

Scale size: 350x350x400mm

Humidification mode: external ultrasonic humidifier (switch control) [optional: intelligent humidity control]

Temperature uniformity: ≤± 0.5 ℃,

Refrigeration power of compressor: 330W

Refrigerants: environmentally friendly fluorine-free R-134a refrigerant

Heater power: 600W

Number of diaphragms: one piece

Timing range: 0-999min

Temperature control range: 5 ℃ - 50 ℃

Temperature uniformity: ± 1 ℃,

Temperature fluctuation: ± 0.1 ℃,

Temperature resolution: ± 0.1 ℃,

Sterilization method: professional ultraviolet germicidal lamp

Control operation form: 1. LED digital display temperature control, P.I.D. microcomputer automatic temperature control.

Power supply: 220 V / 50 Hz


 

3、 Instructions for use:

1. Turn on the power, close the power switch, the whole machine is powered on, and the power indicator in the switch is on. If you need UV sterilization, please turn on the UV switch and turn off the UV lamp when you don't need it.

2. Please refer to the manual of intelligent digital temperature controller for temperature control setting. Keep dry and ventilated. All switches on the front panel should be in non working state. The experimental samples placed in the incubator should be kept at a proper distance. It is conducive to the convection circulation of cold and hot air. Turn on the power supply, turn on the switch, and the indicator light is on; select the culture temperature.

3. If lighting is needed, turn on the lighting switch and set the temperature. Press the set key, and the data in the lower row will flash, which means entering the setting state. Press the arrow down key and the set value will decrease. After pressing the set key to the required setting temperature, it means that the working state is entered. The temperature setting is completed and the setting value is increased. Pressing the arrow up key is the same.

4. The incubator has the function of power-off protection and a minute and a half delay function. It takes about one minute to restart the compressor after shutdown



4、 Maintenance of mould incubator:

1. The shell of incubator should be grounded reliably

2. The incubator should be placed in a cool, dry, well ventilated place away from heat sources and sunlight, and placed stably to prevent vibration and noise.

3. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 10 cm, the side of the box should have a space of 5 cm, and the top of the box should have a space of 30 cm.

4. The incubator should avoid collision, shaking and vibration during handling, repair and maintenance, and the maximum gradient should be less than 45 degrees.

5. If the instrument doesn't work suddenly, please check whether the fuse (behind the box) is burnt out and check the power supply. 6. The difference between the temperature in the incubator and the ambient temperature should not be greater than 25 ℃


 

 

Changzhou Runhua Electric Appliance Co., Ltd

Address: No.26, Jinsheng East Road, Jintan City, Jiangsu Province

Tel: 0519-82305398

Fax: 0519-82306938

website: www.runhuayiqi.com

Case: 394514722@qq.com


}
Video

    No video highlights...
Photo
No photo collection...
Customer
Flash Demo

Project
No related case...
Related articles
Related products
Consulting

Copyright 2020 Changzhou Runhua Electric Co., Ltd. Copyright Design by:EastNet [Manage]
Top