0086-519-82305398
Overview:1. Scope of use The light incubator is the most ideal equipment for cell tissue growth, seed germination, seedling raising, feeding insects and small animals, and microorganism cultivation and preservation. It is especially suitable for the production…
Consulting1. Scope of use
The light incubator is the most ideal equipment for cell tissue growth, seed germination, seedling raising, feeding insects and small animals, and microorganism cultivation and preservation. It is especially suitable for the production and scientific research of thermostatic and humid experiments in the fields of bioengineering, medical research, agricultural science, aquaculture, animal husbandry, etc.
2. RH-250D Technical Indicators:
Work capacity: 250 liters
Temperature control range: 5-50 C without light; 10-50 C with light;
Illumination: 2500-9000LX [light program controlled]
Smart LCD display: six-level adjustable light
Temperature fluctuation: when heated < (+) 1 C, when refrigerated < (+) 1.5 C;
Temperature resolution: 0.1 C
Uniformity: +0.5 C
Timing range: 1-9999min
Power supply: 220V +10% 50Hz;
Carrier Shelf: 3 pieces
Operating size: 470*510*1160
Shape size: 660*610*1770
3. Overview of the structure
he outer shell of this product box is made of high quality steel plate, the surface is sprayed with plastic, the studio is made of stainless steel plate, which is resistant to corrosion. The outer door has an observation window and double-layer hollow glass door, which makes it easy to observe the experiment inside the box.
The intelligent micro-computer controller is easy to operate and can set a variety of parameters (time, temperature, humidity) to simulate the natural environment. The axial flow fan is used to force convection in the box to improve the uniform performance of temperature and humidity in the box.
4. Notes
1. It is forbidden to turn the product upside down or tilt it more than 45 degrees during handling.
2. Do not set temperature and illumination frequently during use to avoid the overload of compressor which will affect the life of compressor.
3. The product is protected by thermal resistance short circuit and circuit break.When the thermal resistance is short-circuited or disconnected, the machine stops heating or cooling automatically.
4. It is not advisable to store too many culture items in the box. Leave gaps in the box to facilitate air circulation in the box.
5. Do not touch or collide with temperature sensor probe and humidity probe in the box to avoid damage and causing uncontrollable.
6. Do not rub the surface with acid-base or other corrosive solutions, but use a dry cloth to clean it.
7. When the equipment is not in use, keep the box dry and cut off the power supply to set all switches at position 0.
8. The product must be operated and maintained by professional personnel. Non-electrical personnel shall not inspect and repair the equipment, and no point inspection shall be allowed.
9. Read the instructions carefully before using the device, and then proceed with various operations.
Welcome to the official website of Runhua Instruments. We are a professional manufacturer of laboratory instruments. Quality and reputation a…
The water bath constant temperature shaker is a biochemical instrument that combines a temperature controllable constant temperature water ba…
Series of biochemical incubators:Features of 150A biochemical incubator: the biochemical incubator has the function of cooling and heating tw…
High-precision stirring water bath produced by Runhua Instruments:1) The general water bath does not have the circulating stirring function. …
Maintenance and maintenance of electric mixer1. Correct use and attention to the maintenance of the instrument to keep it in good working con…
What work should be done in the use of digital display stainless steel electric heating plate? Digital display stainless steel electric h…