Overview:1、 Brief introduction of the instrument: At present, hospitals at all levels generally use vacuum blood collection vessels when collecting blood samples. After completing blood centrifugation, the tube cap must be removed manually, which is inefficie…

Consulting

1、 Brief introduction of the instrument:

  At present, hospitals at all levels generally use vacuum blood collection vessels when collecting blood samples. After completing blood centrifugation, the tube cap must be removed manually, which is inefficient. The vibration generated during decapitation may lead to blood remixing and increase the risk of bacterial infection. The automatic decapitation centrifuge developed by Runhua company has a remarkable effect on decapping, which completely solves the vacuum collection that has plagued hospitals at all levels for many years It is an ideal testing instrument for hospitals.


  The automatic decapitation centrifuge is a centrifugal separation experimental instrument, which uses the rotating centrifugal force to separate the substances in the solution by layers. The instrument adopts stepless speed regulation and automatic balance device, which has the advantages of stable operation, small volume, beautiful appearance, low temperature rise, high efficiency and wide applicability. Special aluminum alloy centrifugal turntable ensures that the centrifugal tube is not easy to be damaged. The instrument is widely used for qualitative analysis of serum and plasma immunity in hospitals, chemical and biochemical laboratories.


Instrument features

1. Microcomputer control, touch panel, LCD imported large LCD display, operation parameters can be automatically memorized, easy to operate and use

2. High torque variable frequency motor drive, imported from Europe, super high speed bearing, stable operation, maintenance free

3. Multi layer damping structure, low vibration, low noise, environmental protection.

4. Steel body, stainless steel centrifugal cavity, key parts imported, easy to clean, durable, no rust.

5. The rotor can be sterilized under high temperature and high pressure to ensure safety.

6. Independent and proprietary air flow guide design, small temperature rise, effective protection of samples.

7. Suitable for: 5ml / 10ml / 15ml / 30ml / 50ml and other capacity adapters to meet various needs

8. Widely used in biopharmaceutical, fine chemical, blood products, medical research.

9. Automatic decapitation, reduce labor intensity, reduce the risk of infection.


2、 Technical parameters:


1. Speed range: maximum speed 3500 rpm

2. Capacity: 5ml × 96

3. Centrifugal force: 2500xg

4. Timing range: 99minh59min59s

5. Dimension: 590 * 430 * 400mm

6. Noise: ≤ 60dB

7. Power supply: 220 V, 50 Hz ± 10%

8. Weight: 55kg

9. Instrument model: TD5B

3、 Safety notice:

  No load operation without rotating head is strictly prohibited: before operation, check whether there is corrosion damage on the rotating head; do not contact the rotating head in operation; do not operate under unbalanced loading condition.

 

4、 Instructions for use

  The working table of the centrifuge should be flat and solid, and the working room should be neat and clean, Open the centrifuge to clean the inner chamber and the rotating head. Put the test tube with the same weight, close the centrifuge cover, adjust the speed control knob to increase the required speed. Before each shutdown, the speed control knob must be set to the minimum position. Set the timer to zero, turn off the power switch and cut off the power supply. Wipe the inner chamber and rotating head, and close the centrifuge cover.


5、 Maintenance

  Do not place any objects on the centrifuge cover. After each experiment, please wipe the inner chamber and rotor to keep it clean, and then change the centrifuge cover to prevent dust and dirt from entering. If the centrifuge is not used for a long time, please open the centrifuge cover for a period of time to dry the inner chamber. After long-term use, the natural wear of the centrifuge is normal. In case of failure or difficulty, the centrifuge should be cleaned Problem, the enterprise will provide you with quality service, lifelong maintenance.

}