Current location:Home > Products > Centrifuge series > Desktop centrifuge
PRODUCTS
  • Mechanical timing knob type stepless speed regulation of 80-2 table type low speed centrifuge

Mechanical timing knob type stepless speed regulation of 80-2 table type low speed centrifuge

Overview:Product introduction Time table centrifuge is a conventional laboratory centrifuge, which is widely used in biology, chemistry, medicine and other scientific research, education and production departments. It is suitable for radioimmunoassay monitorin…

Consulting

Product introduction

  Time table centrifuge is a conventional laboratory centrifuge, which is widely used in biology, chemistry, medicine and other scientific research, education and production departments. It is suitable for radioimmunoassay monitoring and separation of cells and large particles. There are two kinds of timed and irregular centrifuges.

 

2、 Main technical parameters

Working power supply: 220V50Hz

Maximum speed: 4000R / min [stepless speed regulation]

Centrifugal force: 2325 × G

Working capacity: 12 × 20ml

Mechanical timing: 0-60 minutes


 

  It is strictly prohibited to run without load without rotating head; check whether there is corrosion damage on the rotating head before operation; do not contact the rotating head in operation; do not operate under unbalanced loading condition.

 

4、 Instructions for use

  The working table of the centrifuge should be flat and solid, the workshop should be clean, dry and well ventilated, and the ambient temperature should be 5-32 ° C. start the centrifuge and wipe the rotating head clean. Load the weighing test tube, close the centrifuge cover, set the centrifugal time according to your own requirements, adjust the speed control knob to increase the required speed, The speed control knob must be set to the minimum position, the power switch should be turned off, the power supply should be cut off, the inner cavity and the rotating head should be wiped, and the centrifuge cover should be closed.

 

5、 Maintenance

  Do not place any objects on the centrifuge cover. After each use, be sure to clean and wipe the inner cavity and rotating head, and cover the centrifugal cover to prevent dust and dirt from entering. If the centrifuge is not used for a long time, the centrifuge cover should be opened for a period of time to dry the inner chamber. After 3000 hours of use, the carbon brush of the motor should be replaced in time to avoid wearing the commutator, Natural wear and tear is a normal phenomenon. In case of failure or difficulty after one year of use, our company will continue to provide high-quality service to assist in handling.


Changzhou Runhua Electric Appliance Co., Ltd
Address: No.26, Jinsheng East Road, Jintan City, Jiangsu Province
Tel: 0519-82305398
Fax: 0519-82306938
Website: www. runhuayiqi.com
Email: 394514722@qq.com


}
Video

    No video highlights...
Photo
No photo collection...
Customer
Flash Demo

Project
No related case...
Related articles
Related products
Consulting

Copyright 2020 Changzhou Runhua Electric Co., Ltd. Copyright Design by:EastNet [Manage]
Top